Rosemount MS Magnetic Sensor MS300 for Slurry Applications, 30 inch (DN750)
Reference Accuracy ±0.25% with optional ±0.15% high accuracy
Line Size 3” (80mm) to 36” (900mm)
Liner Material Options PTFE, PFA+, Polyurethane, Neoprene, Linatex Natural Rubber, Adiprene
Electrode Options Material: 316L Stainless Steel, Nickel Alloy, Platinum, Tantalum, Titanium
Type: Standard, Bullet, Flat
Optional 3rd electrode for grounding
Flange Options Connections: Slip-on, weld-neck, grooved coupling
Ratings: ASME Class 150 – 2500, EN-1092 PN10 – PN40, AS2129 Table D and E, JIS B2220 10K – 40K, AS4087 PN16 – PN35
Materials: Carbon Steel, 304/304L SST, 316/316L SST
Additional flange ratings and materials may be available upon request
Approvals
Ordinary Locations ATEX Non-Sparking; Dust
Class I Div 2; Dust
Class I Zone 2; Dust
IECEx Non-Sparking; Dust
ATEX increased safety; Dust
Class I Zone1; Dust
IECEx Increase Safety; Dust
Higher level approvals include Intrinsically Safe Electrodes
Enclosure Type Type 4X
IP66, IP67, IP68, IP81
Compatibility
8782 Slurry Transmitter For optimum performance in high noise applications
8712EM and 8732EM (remote mount only) for standard magnetic flow meter performance
Rosemount 8732E Field Mount Magnetic Flow Meter Transmitters
Sensor compatibility Compatible with Rosemount 8705, 8711, and 8721 sensors. Compatible with AC and DC powered sensors of other manufacturers.
Transmitter coil drive current 500mA
Flow rate range Capable of processing signals from fluids that are traveling between 0.04 and 39 ft/s (0.01 to 12 m/s) for both forward and reverse flow in all sensor sizes. Full scale continuously adjustable between –39 and 39 ft/s (–12 to 12 m/s).
Conductivity limits Process liquid must have a conductivity of 5 microSiemens/cm (5 micromhos/cm) or greater.
Power supply 90 – 250VAC, 50/60Hz or 12 – 42VDC
Line power fuses 90-250VAC systems

  • 1A, 250V, I2t, 1.5 A2s Rating, Fast Acting
  • Bussman AGC-1, Littlefuse 31201.5HXP

12-42VDC systems

  • 3 Amp, 250V, I2t, 14 A2s Rating, Fast Acting
  • Bel Fuse 3AG 3-R, Littelfuse 312003P, Schurter 0034.5135

15W maximum – DC

40VA maximum – AC

Switch-on current AC: Maximum 35.7A (< 5ms) at 250VAC

DC: Maximum 42A (< 5ms) at 42VDC